Skip menuCalibrator-meter of harmonic distortions SK6-20.
Purpose:
- Calibrator-meter of harmonic distortions is designed for
- THD measurement;
- measurements of first harmonic frequency and amplitude;
- generation of sinusoidal signal with given standardized THD value with external generator-calibrator of harmonic signals SK6-122-01 attached (or Agilent Technologies arbitrary waveform generator 33220A), which is controlled through USB;
- amplitude-frequency response and phase-frequency response calibration of quadripoles.
- SK6-20 can be used as second grade working standard of THD value.
Technical specifications:
- Basic specifications in THD measurement mode.
- First harmonic frequency range in THD measurement mode 0,01 ... 200 kHz
- THD values measurement range 0,001 100 %
- First harmonic amplitude measurement range (under condition that RMS value of input signal is less than 1.27V) in THD measurement mode 1 ... 1,8 V
- Maximum absolute error of THD value measurement depending on first harmonic frequency is given in following table:
- Basic specifications in amplitude and frequency of first harmonic measurement mode.
- First harmonic frequency range 0,01 200 kHz
- First harmonic amplitude range (under condition that RMS value of input signal is less than 1.27V) 0,01 ... 2,00 V
- Limits of first harmonic frequence measurement relative error 0,006 %
- Limits of first harmonic amplitude measurement relative error:
     -in range form 10 mV to 1 V: 1,8 %
     -in range from 1 V to 2 V: 0,6 % - Basic characteristics in generating signal with given THD value mode (with external generator-calibrator of harmonic signals SK6-122-01 attached)
- First harmonic frequency range in generation mode 0,01 200 kHz
- THD value generation range 0,001 100 % with minimum step of 0,001 %
- Limits of THD value generation absolute error depending on first harmonic frequency are given in following table:
First harmonic frequency range |
Limits of THD measurement absolute error, % |
(10,0 199,9) Hz | (0,03 × THD + 0,006 %) |
(0,2 20,0) kHz | (0,02 × THD + 0,005 %) |
(20,001 200) kHz | (0,04 × THD + 0,006 %) |
First harmonic frequency range |
Limits of absolute THD value generation error, % |
(10,0 199,9) Hz | (0,03 × THD + 0,006 %) |
(0,2 20,0) kHz | (0,02 × THD + 0,005 %) |
(20,001 200) kHz | (0,04 × THD + 0,006 %) |